Products

Featured products

  • Diffractometer
    Diffractometer
    1. The accuracy of Diffractometer is high. 2. The application range of Diffractometer is wide. 3. Diffractometer is easy to operate, convenient and efficient.
    More
  • Single Crystal XRD
    Single Crystal XRD
    1.The single crystal machine adopts PLC control technology. 2.Modular design, accessories plug and play. 3.Electronic lead door interlocking equipment with double protection. 4.Single crystal X-ray tube: a variety of targets can be selected, such as Cu, Mo,etc. 5. Single crystal adopts four-circle concentric technology to ensure that the center of no goniometer remains unchanged.
    More
  • A 2D X-ray Diffractometer
    A 2D X-ray Diffractometer
    Advantages: Continuously adjustable X-ray penetration depth Capability to observe the distribution of crystal planes with different orientations Analysis of orientation distribution in samples such as fibers, thin films, and powders Examination of structural characteristics like lattice distortion and crystallite size
    More
  • X-ray Diffraction Residual Stress Analyzer
    X-ray Diffraction Residual Stress Analyzer
    TD-RSD XRD Residual Stress Analyzer ±7MPa accuracy, dual-detector innovation, full-auto operation & powerful software. Optimize processes, enhance product reliability. Compact, efficient, leading solution.
    More

Contact us

X-ray diffractometer operation tips

2023-08-05

Today, let's share some knowledge of X-ray diffractometer operation, mainly in the following aspects.


一、 Turn on


1. Check whether the XRD instrument is normal and whether the cooling water circulating machine is leaking;

2. Turn on the wall power supply, cooling water circulating machine, and then turn on the XRD system power supply (orange on the right close) and preheat for 10-15 minutes.


二、Sample preparation and testing

       1. Prepare the sample and put it into the equipment sample fixture;


       2. Turn on the computer and open the desktop shortcut of the software;


       3. Click successively:

 "Control" - "Manual control" - "Online" - "Close";

 "Collection" - "Scan parameter Settings", parameter Settings;

 Scanning mode - continuous; Drive mode one double shaft;

 Starting Angle - not less than 1 degree; Termination Angle - No more than 140 degrees.

 Starting Angle - not less than 1 degree; Termination Angle - No more than 140 degrees.

 The default pipe pressure is 30KV, the pipe flow is 20MA, if you need to modify, please contact the responsible personnel.

    

     4. Click "Apply" to start the XRD pattern test;

    

     5. When "End of work" is displayed, click "OK" and then "Save spectrum" to save the data:

   

     6. When multiple samples are continuously measured, the sample can be changed at this time and then tested:

 

 If you do not need to change the parameters to continue the test, click "Start scanning" to continue the test:


To change scan parameters, click Scan Parameters to reset the parameters.


三、 Turn off



1. After the test, turn off the software and the power supply of the XRD system successively;



2. After 15 minutes, turn off the cooling water circulating machine and wall power supply.